Gas Detectors & Bump Testing

Many deadly gases have no color or smell, and your nose can’t be trusted even with the ones that do. A portable gas detector is the only way to know whether the air in a confined space, trench, tank or plant is safe to breathe. But a detector only protects you if it’s working, tested and used the right way.

Photo: 4-gas monitor clipped near a worker’s collar, next to a bump test station

What a 4-gas monitor measures

The most common portable monitor checks four things:

ReadingWhy it mattersTypical alarm settings
Oxygen (O2)Normal air is 20.9% oxygen. Low oxygen causes confusion, collapse and death. High oxygen makes things burn explosively.Low alarm 19.5%; high alarm 23.5%
Combustible gas (LEL)Shows how close the air is to the lower explosive limit of flammable gases like methane, propane and gasoline vaporLow alarm usually 10% of the LEL
Carbon monoxide (CO)From engines, generators, heaters and fires. Odorless and toxic.Often 35 ppm (some set 25 ppm)
Hydrogen sulfide (H2S)From sewers, manure, oil and gas, and pulp mills. Kills your sense of smell at high levels.Often 10 ppm (some set 1 to 5 ppm)

Alarm settings are set by your employer, based on regulations, exposure limits and the manufacturer. Know what your monitor’s alarms mean before you use it.

Other types of detectors

Bump testing vs. calibration

Bump testCalibration
What it isA quick check that exposes the sensors to test gas to make sure they respond and the alarms go offAdjusting the sensors so their readings match a known concentration of test gas
How oftenBefore each day’s use (industry recommendation)As the manufacturer requires, and whenever a bump test fails
TakesAbout a minuteA few minutes

Sensors can fail without warning. They can be damaged by drops, water, extreme temperatures, or poisoned by silicones, lead and other chemicals. A monitor that hasn’t been bump tested may show “all clear” in deadly air. The International Safety Equipment Association recommends a bump test before each day’s use.

Before you use it

  1. Check that the battery is charged and the monitor turns on and passes its self-test.
  2. Make sure calibration and bump test dates are current.
  3. Zero it in fresh, clean air, away from exhaust and the work area.
  4. Bump test it with the test gas, and confirm every sensor and alarm (sound, light and vibration) works.
  5. Check the filter, inlet and sampling tubing for dirt, water or cracks.
  6. Record the test, if your program requires it.

Testing a confined space or trench

See Confined Spaces and Access, Egress, Spoils & Water.

Wearing a personal monitor

When it alarms

  1. Leave the area immediately and go to fresh air. Don’t try to finish what you were doing.
  2. Warn others and keep them out.
  3. Tell your supervisor. Don’t go back in until the source is found, ventilated or controlled, and the air tests safe.
  4. Never assume it’s a false alarm. Investigate before re-entry.

Danger: Never enter a space to rescue someone who collapsed unless you are a trained rescuer with the right equipment, including supplied air. In many confined space deaths, more than half of the victims were would-be rescuers.

Things that fool gas detectors

Daily checklist

Video: how to bump test a 4-gas monitor and test a confined space at three levels

Standards & references

Related: Confined Spaces · Hydrogen Sulfide · Carbon Monoxide · Hot Work Permits

Last updated: